When setting voltages it should be understood that all photomultiplier tubes (PMTs) operate optimally within certain ranges. The factors influencing detector performance can include, but are not limited to: systemic electronic noise, detector quantum efficiency, optical filter quality, laser alignment, detection wavelengths, and the voltage applied to the PMTs.
Since all cells are variable with respect to their background autofluorescence, there is no “one size fits all” group of instrument settings that can be universally applied to the PMTs on a particular flow cytometer. Moreover, instrument settings for negative populations should not be arbitrarily assigned to low fluorescence values (as was done in older systems), but through a more methodical process; it is known that insufficient gains (voltage) can decrease resolutions between negative and positive populations, whereas high gain values can push very bright populations off-scale making them less easy to quantify.
When these procedures are followed, cytometer performance will be traceable over the long term, as is often necessary when repeating the same experiment many times or in clinical trials. In order to limit the influence of external factors when establishing instrument settings, the following guidelines should be followed:
Table 1: Example Baseline Data for BD Cytometer Setup and Tracking Beads
The minimum PMT voltage setting for each detector is determined so that the resulting rSD of the unstained cells is approximately 2.5x the rSD of the electronic noise (2.5 rSDen).
Purpose: to ensure that electronic noise does not interfere with measurements with dim fluorescence and/or negative events and that you are aware of these limits should it be necessary to decrease the voltages lower than these values in Rule 2 (below).
The Short Way:
The Long Way:
The Short Way:
The Long Way:
If the positives are so far up the scale using the minimum acceptable voltage that they are in the nonlinear range or going off scale, the voltage must be lowered to properly view the positive data regardless of where the negatives reside.
Placing negative cells at 2.5x rSDen does not always result in the best resolution between the negative and positive fluorescence. For example, at the minimum voltage, the separation between negative and positives can be reduced when compared to voltages closer to the CST values; this can consequently have a negative impact on the detection of weakly expressed antigens. Alternatively, when too much voltage is applied, negative populations can spread, especially in channels where there is little to no detectable autofluorescence (red emissions), therefore it is important to optimize the resolution of all stains, paying particular attention to dim fluorochromes and weakly expressed markers, to ensure their correct identification by characterizing the Stain Index over a wider range of voltages.
Figure 1: The effective brightness of a reagent depends on the difference (D) between the positive (red) and the negative (blue, green) populations and the spread of the negative population (W). The stain index is a useful metric for normalized signal over background [1].
The following method will help to minimize the spread of the negatives while optimizing the resolution of the stains for each detector.
Procedure: Acquire data for the single stained control cells, over a wider range of voltages to determine where the best resolution is obtained. Use this voltage instead of the rSDen as your starting voltage.
BD FACSDIVA based “Application Settings” are used to standardize your experiments across time and between instruments. If the CST program has changed the system voltages to separate the dim bead population apart from noise, the software will automatically adjust your specific application settings, to a proportional degree, so that populations will appear in the same position on the scale each time the CST values are updated (daily).
Procedure:
RCP Manufacturer: Spherotech, Catalogue No: URCP-38-2K
There are six peaks of fluorescence in these beads; choose the peak that best suits your data. Use this same peak to match your MFIs each day. Recommended for experiments with many red detection parameters and UV excitations.
Manufacturer: Spherotech, Catalogue No: RFP- 30-5A
These beads have poor red/far-red fluorescence and do not work for UV excited fluorochromes. Best for FITC/PE/PE-Cy5 and violet excitations
Rainbow beads may be purchased directly from the manufacturer (spherotech.com) or Cedarlane Laboratories.
[1] Maeker, H.T., & Trotter, J. Selecting reagents for multi-colour flow cytometry with BD™ LSRII and BD FACS Canto™ systems. Nature Methods 5 (2008), Application Note.
[2] Maecker, H.T., Frey, T., Nomura, L.E. & Trotter, J. Selecting fluorochrome conjugates for maximum sensitivity. Cytomety A 62, 169–173 (2004).
[3] Baumgarth, N. & Roederer, M. A practical approach to multicolor flow cytometry for immunophenotyping. J. Immunol. Methods 243, 77–97 (2000).
[4] Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software – Technical Bulletin
Lorem ipsum dolor sit amet, consectetur adipiscing elit, sed do eiusmod tempor incididunt ut labore et dolore magna aliqua.
7226-1 King’s College Circle
Medical Sciences Building University of Toronto
416-976-6524